< 所有产品分类
张力计

TI-MINIMAX螺栓张力计
- 产地:中国
- 上架时间:12-02-23 14:39
- 产品描述
- 技术参数
- 技术文档
- 相关产品
- 如何购买
产品描述:
The TI-MINIMAX can measure the elongation very accurately in fasteners of virtually any material from 1/2 inch to 4 feet in length. The measurement is achieved by determining the change in the transit time, of an ultrasonic shock wave along the length of the fastener as the fastener is tightened by any method. The on-board microcomputer automatically interprets this time measurement to display the time(nanoseconds), elongation, load, stress, or %strain from stretching a fastener.
|
|
技术参数:
Measuring Range
|
From 1” to 48” (25.4mm to 122cm) bolts
|
Time
|
Nanoseconds
|
Elongation
|
Change in length (inches/millimeters)
|
Load
|
Force load applied (pounds kIP or kN Killonewtons)
|
Stress
|
Force for unit area stress applied (pounds kPSI or megapascals MPa)
|
Strain %
|
percentage strain applied (inches per inch or millimeters per millimeter)
|
Resolution
|
+/- .00001 inches (0.0001 mm)
|
Velocity Range
|
.0492 to .3936 in/ms
1250 to 9999 meters/sec |
Calibration
|
Fixed, Single, and Two point zero calibration options.
Select bolt material types from a preset or custom list.
|
Units
|
English & Metric / Fahrenheit & Centigrade
|
Measuring Modes
|
Pulse-Echo (standard)
Pulse-Echo w/Gate (fine adjust) |
Pulser
|
Square wave pulser with adjustable pulse width (spike, thin, wide)
|
Receiver
|
Manual or Auto Set gain control with 40dB range
|
Timing
|
10 bit 250 MHz digitizer
|
Display Modes
|
A-Scan - Rectified +/- (half wave view), or RF (full waveform view)
|
Alarm Limits
|
Set hi and lo tolerances with audible beeper, viewable scan bar, and visual LEDs
|
Data Logger (Internal)
|
Total of 8,000 readings, multiple bolt groups. Stores both waveform views, nanoseconds, elongation, load, stress, and strain for each reading
|
Memory
|
16 megabit non-volatile ram
|
Transducer
|
Single Element (1 to 10 MHz & 1/8” to 1” diameters)
|
Transducer Cable Lenght
|
10 foot (custom lengths available, inquire)
|
Temperature Probe
|
for automatic temperature compensation
|
Setup Memory
|
64 custom user defined setups. Factory setups can also be edited by the user
|
Gate
|
Gate used to fine adjust where the detection point occurs
|
Auto Set
|
Locates the detection signal, optimizes the gain setting, and adjusts the overall display to show the waveform and detection point automatically
|
Field Calibration
|
Vector & linear regression. Includes Factory calibration traceable to national standards
|
Power
|
Three 1.5V alkaline or 1.2V NiCad AA cells.
Typically operates for 150 hours on alkaline and 100 hours on NiCad (charger not included.)
|
Dimensions
|
Width (2.5in/63.5 mm)
Height (6.5 in/165 mm) Depth (1.24 in/31.5 mm) |
Weight
|
13.5 ounces (with batteries)
|
Operating Temperature
|
14 to 140F (-10C to 60C)
|